Accurate pH measurement is vital across research, manufacturing and environmental testing. Yet, achieving and maintaining accuracy can be challenging due to temperature effects, electrode maintenance requirements and sample variability. Reliable methods are essential to ensure data integrity and reproducibility.
This guide highlights the core principles of pH theory, as well as electrode selection, calibration and troubleshooting tips, to help professionals refine measurement accuracy and consistency.
Download this guide to discover:
- Key factors that impact pH accuracy and stability
- How to select and maintain electrodes for optimal performance
- Proven calibration and measurement techniques for dependable results
pH measurement handbook
thermo scientific
pH theory 4
pH electrode theory 4
pH versus temperature theory 5
pH electrode selection guide 6
Thermo Scientific pH electrode families 7
pH electrode features 8
pH electrode recommendations by sample 11
pH electrode catalog numbers by sample type 12
pH electrode summary table 14
pH electrode operation guide 16
pH electrode preparation 16
Thermo Scientific pH buffers and storage solutions 16
pH electrode calibration 17
Calibration recommendations 17
Three-buffer calibration procedure 17
pH electrode measurements 19
Measurement recommendations 19
Sample requirements 20
pH measurement procedure 20
pH electrode maintenance 20
Filling and draining a refillable pH electrode 20
Contents
pH electrode storage 22
Short-term electrode storage (up to one week) 22
Long-term electrode storage (more than one week) 22
pH electrode troubleshooting guide 23
General troubleshooting procedures 23
pH electrode interferences 23
pH electrode cleaning procedures 24
Role of solution quality in pH measurement 25
Recommended shelf life for buffers and solutions 25
Common pH sample issues and remedies 26
pH measurements in TRIS buffer, protein and sulfide samples 26
pH measurements in solid samples and surface pH measurements 26
pH measurements in pure water samples 26
pH measurements in sludges, suspensions, colloids, slurries and viscous samples 27
pH measurements in extreme pH or high salt content samples 27
pH measurements in nonaqueous samples 28
4
The term pH is derived from a combination of “p” for the word power and “H” for the symbol of the element hydrogen. Together the meaning is the power of hydrogen. pH serves as a convenient way to compare the relative acidity or alkalinity of a solution at a given temperature. A pH of 7 describes a neutral solution because the activities of hydrogen and hydroxide ions are equal. When the pH is below 7, the solution is described as acidic because the activity of hydrogen ion is greater than that of hydroxide ion. A solution is more acidic as the hydrogen ion activity increases and the pH value decreases. Conversely, when the pH is above 7, the solution is described as basic (or alkaline) because the activity of hydroxide ion is greater than that of hydrogen ion.
pH electrode theory
The activity of the hydrogen ion in solution is measured with a pH measuring system consisting of a glass electrode, a reference electrode, and a pH meter. When a pH sensing glass electrode comes in contact with a sample, a potential develops across the sensing membrane and that membrane potential varies with pH. This sensing electrode is known as a half-cell and a reference electrode provides a second, unvarying potential to quantitatively compare against. Combination pH electrodes are composed of a sensing electrode with the reference electrode built into the same electrode body. Combination electrodes provide the same selectivity and response as a half-cell system, but offer the convenience of working and maintaining only one electrode. A meter serves as the readout device and calculates the difference between the reference electrode and sensing electrode potentials in units of millivolts. The millivolt response is used to calculate the pH reading as displayed on the meter. Known pH standards (buffers) traceable to NIST standards are referenced to calibrate the system prior to any readings.
pH theory
Electrode behavior is described by the Nernst equation: E = Eo + (2.3 RT/nF) log aH+
E is the measured potential from the sensing electrode, Eo is the fixed potential of the reference electrode, (2.3 RT/nF) is the Nernst factor and log aH+ is the pH. The Nernst factor, 2.3 RT/nF, includes the Gas Law constant (R), Faraday’s constant (F), the temperature in degrees Kelvin (T) and the charge of the ion (n). For pH, where n = 1, the Nernst factor is 2.3 RT/F. Since R and F are constants, the factor and therefore electrode behavior is dependent on temperature.
The electrode slope is a measure of the electrode response to the ion being detected and is equivalent to the Nernst factor. When the temperature is 25 °C, the theoretical Nernst slope is 59.16 mV/pH unit. Thermo Scientific pH meters display the slope as a percentage of the theoretical value. For example, a 98.5 % slope is equivalent to a slope of 58.27 mV/pH unit for a two-point calibration. The pH meter detects the pH sensing bulb signal, reference signal and temperature signal and uses these values to calculate the pH using the Nernst equation. Thermo Scientific™ pH meters contain pH versus temperature values for commonly used buffers. This allows the meter to recognize a particular pH buffer and calibrate with the correct buffer value at the measured temperature.
5
The most common cause of error in pH measurements is temperature. Temperature variations can influence pH for the following reasons:
•
The electrode slope will change with variations in temperature.
•
Buffer and sample pH values will change with temperature.
•
Measurement drift can occur when the internal elements of the pH and reference electrodes are reaching thermal equilibrium after a temperature change.
•
When the pH electrode and temperature probe are placed into a sample that varies significantly in temperature, the measurements can drift because the temperature response of the pH electrode and temperature probe may not be similar and the sample may not have a uniform temperature, so the pH electrode and temperature probe are responding to different environments.
Electrode slope changes can be compensated for by using an automatic temperature compensation (ATC) probe or a Thermo Scientific™ Orion™ Triode™ electrode, which has an ATC probe incorporated into the body of the electrode. The pH meters calculate the electrode slope based on the measured temperature of the pH buffers. The meter will automatically adjust the pH buffer value to the actual pH of the buffer at the measured temperature.
The pH values of buffers and samples will change with variations in temperature because of their temperature
pH versus temperature theorydependent chemical equilibria. The pH electrode should be calibrated with buffers that have known pH values at different temperatures.
Each sample is unique with thier own undefined pH vs. temperature effect. Therefore, a pH meter cannot correct a sample pH to a reference temperature. Whenever possible, calibrate your system near the sample temperature and always record sample temperature with each reading.
pH value changes by temperature variation
Nominal pH Value
at 25 °C
0°C
5°C
10°C
20°C
30°C
40°C
50°C
60°C
70°C
80°C
90°C
1.68
1.67
1.67
1.67
1.68
1.68
1.69
1.71
1.72
1.74
1.77
1.79
4.01
4.00
4.00
4.00
4.00
4.02
4.03
4.06
4.09
4.12
4.16
4.21
6.86
6.98
6.95
6.92
6.87
6.85
6.84
6.83
6.84
6.85
6.86
6.88
7.00
7.11
7.08
7.06
7.01
6.98
6.97
6.96
6.97
7.00
7.03
7.08
9.18
9.46
9.40
9.33
9.23
9.14
9.07
9.01
8.96
8.92
8.89
8.85
10.01
10.32
10.25
10.18
10.06
9.97
9.89
9.83
9.79
9.78
9.78
9.80
12.46
13.47
13.24
13.03
12.64
12.29
11.99
11.73
11.50
11.30
11.13
10.98
Electrode Potential (mV)
714pH100°C (74 mV/pH Unit)50°C (64 mV/pH Unit)0°C (54 mV/pH Unit)Isopotential Point5000-5000
6
The types of available electrodes, desired electrode features and the compatibility of the electrode with the sample are all determining factors in selecting the right pH electrode. The following section provides the information needed to select the best pH electrode for your applications. The pH electrodes are available with a glass or epoxy body and a variety of electrode body styles, fill types, references, junctions and connectors that suit your application. Other factors to consider when selecting a pH electrode include:
•
The reference junction should provide a stable and reproducible reading under a wide variety of sample conditions. Thermo Scientific™ electrodes include a large selection of reference junctions and filling solutions that are designed to provide fast and reliable readings in all types of samples. Electrodes with a Thermo Scientific™ Sure-Flow™ reference junction can be used in a variety of samples, including difficult samples such as dirt, colloids, sludge and viscous material.
•
The electrode filling solution should not interfere with the electrode measurements. Electrodes with a double junction reference prevent silver ions from coming in contact with the sample, which is key when measuring samples that contain silver-binding agents such as TRIS buffer, proteins or sulfides. Refillable Thermo Scientific™ Orion™ ROSS Ultra™ and Thermo Scientific™ Orion™ ROSS™ electrodes do not contain any silver and allow the operator to modify the outer filling solution. This is useful when the regular filling solution contains ions that interfere or react with the sample that is being measured.
•
The filling solution should flow freely with no fouling or clogging of the junction by the sample. The Sure-Flow reference junction on many Thermo Scientific pH
pH electrode selection guideElectrodes improves the electrode performance by allowing a constant flow of filling solution into the sample. The enhanced flow rate produces stable reference potentials for faster response and better stability. In addition, pressing on the electrode cap flushes the reference, which allows for easy cleaning and refilling of the electrode. Thermo Scientific™ AquaPro™ pH electrodes incorporate a high-performance polymer that isolates the silver internal reference from the sample. These low maintenance pH electrodes offer a rapid response due to their open junction design, even in dirty or viscous samples.
Fill hole
InternalreferenceReferencefilling solutionchamberReference junctionpH-sensingbulbCap
Figure 1 – Typical pH Electrode Components
7
ROSS Ultra Premium pH Electrodes
The best of the best. The fastest response, highest accuracy and best reproducibility that you can get with a pH electrode – plus an 18 month (gel-filled) or 24 month (liquid-filled) warranty.
ROSS Ultra TriodeTM pH/ATC Electrodes
These electrodes include a built-in temperature sensor to measure both pH and temperature with one electrode, minimizing the amount of required sample.
ROSS Premium pH Electrodes
Precise pH measurements with unmatched stability, reliability and rapid response regardless of temperature or sample composition.
ROSS Micro pH Electrode
Measure samples as small as 15 μL in 384 well plates. Use with the Thermo Scientific™ Orion™ Versa Star™ LogR™ Meter to take pH and temperature measurements utilizing LogR technology.
No Cal pH Electrodes
The world’s first pH electrode that does not require calibration. No Cal pH electrodes maintain their calibration to 0.1 pH units for one year when properly stored and maintained.
AquaProTM Professional pH Electrodes
Low maintenance, clog resistant double junction electrodes with an extended life and fast response.
Thermo Scientific pH electrode families
Standard Ag/AgCl pH Electrodes
A large variety of quality electrodes for a wide range of applications. Includes specialty pH electrodes for unique or challenging pH measurements.
Green pH Electrodes
These environmentally friendly electrodes contain no lead, mercury or other hazardous substance, which allows for easy and hassle-free electrode disposal and meets all the requirements of Restriction of Hazardous Substances (RoHS).
Micro Ag/AgCl pH Electrodes
These pH electrodes measure samples as small as 0.5 μL in containers as small as 384 well plates.
Double Junction pH Electrodes
Isolated Ag/AgCl reference system that prevents silver from coming in contact with the sample and is great for measuring TRIS buffer, sulfide and protein samples.
Economy pH Electrodes
Good performance, value-priced, durable and low maintenance gel-filled pH electrodes.
ATC Probes
Automatic temperature compensation probes measure sample temperatures and adjust pH measurements by correcting the electrode slope according to the measured temperature.
8
pH electrode features
Standard size
12 mm electrode diameter for use in a wide variety of sample sizes
Semi-micro
6 to 8 mm electrode diameter for sample sizes down to 200 μL
Micro
1 to 5 mm electrode diameter for samples as small as 0.5 μL and containers as small as 384 well plates
Rugged bulb
When glass is required, the extra durable pH bulb prevents breakage
Spear tip
For piercing solid or semi-solid samples and measuring small volume samples
Flat surface tip
For measuring surfaces of solid or gel samples and measuring small volume samples
pH electrode styles
Glass body
•
Compatible with virtually any sample, including solvents
pH electrode body materials
Epoxy body
•
Extremely durable and rugged to prevent breakage
Sure-Flow pH electrode, sleeve and laser drilled hole (open)
•
Best junction for
dirty, difficult samples
•
Junction is clog-free
and easy to clean
•
Ideal for thick or viscous samples, compatible with all sample types
Wick and glass fiber
•
Good junction for routine lab or field use
•
Junction used with rugged epoxy electrodes
•
Ideal for aqueous samples
pH electrode junctions
Ceramic and glass capillary
•
Better junction for
routine lab or field use
•
Junction is high quality and durable
•
Ideal for most applications
and samples
9
Refillable
•
Easy maintenance, filling solution required
•
Periodic filling and draining needed
•
Long expected life span
•
0.01 to 0.02 pH precision
•
Best response time
•
1 or 2 year warranty (6 months
for Ag/AgCl micro electrodes)
Gel filled (sealed)
•
Low maintenance, no filling solution needed
•
Sealed reference
•
Easy to use, general purpose
•
0.05 to 0.1 pH precision
•
Good response time
•
6 month warranty
pH electrode fill types
Polymer filled (sealed)
•
Low maintenance, no filling solution needed
•
Sealed reference
•
Easy to use, ideal for most applications
•
0.02 pH precision
•
Better response time
•
1 year warranty
ROSS reference
•
0.01 pH precision
•
Best measurement response time
•
Ideal for TRIS, protein and sulfide samples
•
Variety of body styles and types
•
Refillable or gel design
•
Best temperature response
Double junction Ag/AgCl reference
•
0.02 pH precision
•
Better measurement response time
•
Ideal for TRIS, protein and sulfide samples
•
Variety of body styles and types
•
Refillable, polymer or gel design
•
Good temperature response
Single junction Ag/AgCl references
•
0.02 to 0.1 pH precision
•
Good measurement response time
•
General purpose for everyday use
•
Variety of body styles and types
•
Refillable or gel design
•
Good temperature response
pH electrode references
10
pH Electrode Connectors
ATC Probe Connectors
Waterproof BNC
Compatible with any pH meter that has a BNC input. Ensures a waterproof and secure connection with the Thermo Scientific™ Orion Star™ A series meters.
MiniDIN
Compatible with the Orion Star A, Orion™ Dual Star™ and Orion Versa Star Pro meters and pH meters that have an 8 pin MiniDIN ATC input.
BNC
Compatible with any pH meter that has a BNC input.
U.S. standard
Compatible with older style meters that have a U.S. standard input.
RCA
Compatible with pH meters that have an RCA input.
Screw cap
Electrode can be connected to a variety of meter inputs using a separate cable.
Pin tip
Compatible with meters that have a reference half-cell input.
ATC probe ordering information
Description
Cat. No.
Epoxy ATC Probe, MiniDIN
927005MD
Glass ATC Probe, MiniDIN
927006MD
Stainless steel ATC Probe, MiniDIN
927007MD
Micro ATC Probe with Stainless Steel
1 mm x 38 mm Tip, MiniDIN
928007MD
Adapter cable ordering information
Description
Cat. No.
LEMO Miniature Meter to
BNC pH Electrode Adapter
090035
BNC Meter to Screw Cap pH Electrode Cable
91CBNC
11
pH electrode
recommendations by sample
pH Electrode recommendations by sample
Sample type
Electrode recommendations
Biological/Pharmaceutical – TRIS buffer, proteins, enzymes
A ROSS or double junction Ag/AgCl reference
(no sample contact with silver)
Education/student use
An epoxy body for added durability
Emulsions – foods, cosmetics, oils
A Sure-Flow or open junction to prevent the electrode from clogging
Emulsions – petroleum products, paint
A glass body that resists damage from the sample and a Sure-Flow or open junction to prevent the electrode from clogging
Flat surfaces – cheese, meat, agar
A flat-surface tip and ROSS or double junction Ag/AgCl reference (no sample contact with silver)
Flat surfaces – paper
A flat-surface tip
General purpose – most sample types
All electrodes are suitable for general purpose measurements
Harsh environments – field or plant use, rugged use
An epoxy body for added durability and polymer or gel filled for easy maintenance
High ionic strength - acids, bases, brines,
pH > 12 or pH < 2
A Sure-Flow or open junction for better contact with the sample and more stable measurements
High temperatures
A ROSS reference for longer life and/or a quick flow junction for fast response
HF samples
A rugged or durable glass bulb and a double junction to protect the reference
Large sample sizes - tall flasks
A long body that fits the container
Low ionic strength - treated effluent, deionized water, distilled water
Refillable for better contact with the sample and more stable measurements
Nonaqueous – solvents, alcohols
A glass body that resists damage from the sample and a Sure-Flow junction for better contact with the sample and more stable measurements
Semi-solids - fruit, meat, cheese
A spear tip for piercing samples and a ROSS or double junction Ag/AgCl reference
Small sample size - micro-titer plates
A small diameter that fits the container
Small sample size – NMR tubes
A small diameter that fits the container
Small sample size – test tubes, small flasks and beakers
A small diameter that fits the container
Small sample size – TRIS
buffer, proteins, sulfides
A small diameter that fits the container and a ROSS or double junction Ag/AgCl reference
Titration
A Sure-Flow or sleeve junction for better contact with the sample and more stable measurements
Viscous liquids - slurries, suspended solids sludges
A Sure-Flow or open junction to prevent the electrode from clogging
Waters – acid rain, boiler feed water, distilled water, rain water, well water
A ROSS or double junction Ag/AgCl reference that is refillable for better contact with the sample
Waters – drinking water, tap water
An epoxy body for added durability
Waters – wastewater, seawater
A ROSS or double junction Ag/AgCl reference and have an epoxy body for added durability
12
ROSS Ultra
ROSS
AquaPro
Standard Ag/AgCl
Green
Micro Ag/AgCl
Double Junction
Economy
Biological/pharmaceutical –
TRIS buffer,
proteins, enzymes
8102BNUWP, 8107BNUMD 8156BNUWP, 8157BNUMD 8302BNUMD, 8107BNURCA, 8157BNURCA, and 8302BNURCA
8102BN,
8104BN
815600, 8165BNWP 8172BNWP
9104APWP
9107APMD
9156APWP
9165BNWP,
9172BNWP
GD9106BNWP GD9156BNWP
9110DJWP
9102DJWP 9156DJWP
Education/student use
8107BNUMD, 8156BNUWP 8157BNUMD, 8107BNURCA, 8157BNURCA
815600,
8165BNWP
9107APMD
9156APWP
9107BNMD, 9156BNWP,
9157BNMD
GD9106BNWP
9156DJWP
9106BNWP
Emulsions –
foods, cosmetics, oils
8165BNWP
8172BNWP
9104APWP
9107APMD
9165BNWP,
9172BNWP
9110DJWP
Emulsions –
petroleum
products, paint
8172BNWP
9104APWP
9172BNWP
Flat surfaces –
cheese, meat, agar
8135BNUWP
8135BN
9135APWP
Flat surfaces –
paper
8135BNUWP
8135BN
9135APWP
913600
General purpose –
most sample types
8102BNUWP, 8107BNUMD 8156BNUWP, 8157BNUMD 8302BNUMD, 8107BNURCA, 8157BNURCA, and 8302BNURCA
8102BN,
8104BN
815600, 8165BNWP 8172BNWP
9104APWP
9107APMD
9156APWP
9102BNWP, 9107BNMD, 9156BNWP,
9157BNMD
GD9106BNWP GS9156BNWP GD9156BNWP
9110DJWP
9102DJWP 9156DJWP
9106BNWP
Harsh environments –
field or plant use, rugged use
8107BNUMD,
8107BNURCA
9107APMD
9156APWP
9107BNMD
GD9106BNWP
9156DJWP
9106BNWP
High ionic strength – acids, bases, brines, pH > 12 or pH < 2
8165BNWP
8172BNWP
9104APWP
9107APMD
9165BNWP
9172BNWP
High temperatures
8157BNUMD, 8102BNUMD, 8107BNUMD
8165BNWP,
8172BNWP
9165BNWP,
9172BNWP
HF samples
8302BNUMD
8172BNWP, 8104BN
9104APWP
Large sample sizes – tall flasks
912600
pH electrode catalog numbers by sample type
13
ROSS Ultra
ROSS
AquaPro
Standard Ag/AgCl
Green
Micro Ag/AgCl
Double Junction
Economy
Low ionic strength –
treated effluent, deionized water, distilled water
8102BNUWP,
8156BNUWP 8157BNUMD, 8302BNUMD, 8157BNURCA
and 8302BNURCA
8102BN, 815600 8165BNWP
8172BNWP
9165BNWP,
9172BNWP
GS9156BNWP GD9156BNWP
9102DJWP
Nonaqueous – solvents, alcohols
8172BNWP
9172BNWP
Semi-solids –
fruit, meat, cheese
8163BNWP
9120APWP
Small sample size – Micro-titer plates
8220BNWP
9810BN
Small sample size –
NMR tubes
9826BN
Small sample size –
test tubes, small flasks and beakers
8103BNUWP, 8115BNUWP
8103BN, 8115BN 8175BNWP
9103APWP
9115APWP
9103BNWP
9810BN,
9826BN
9110DJWP
911600,
912600
Small sample size –
TRIS buffer, proteins, sulfides
8103BNUWP, 8115BNUWP
8103BN, 8115BN 8175BNWP
9103APWP
9115APWP
9110DJWP
Titration
8162SC, 8172BNWP, 8165BNWP
Viscous liquids – slurries, suspended solids sludges
8165BNWP 8172BNWP
9104APWP
9107APMD
9165BNWP, 9172BNWP
Waters –
acid rain, boiler feed water, distilled water, rain water, well water
8102BNUWP, 8156BNUWP 8157BNUMD, 8302BNUMD, 8157BNURCA and 8302BNURCA
8102BN, 815600 8165BNWP 8172BNWP
9165BNWP,
9172BNWP
GS9156BNWP GD9156BNWP
9102DJWP
Waters –
drinking water,
tap water
8156BNUWP, 8157BNUMD, 8157BNURCA
815600, 8165BNWP
9107APMD
9156APWP
9156BNWP, 9165BNWP
GD9106BNWP
9156DJWP
9106BNWP
Waters –
wastewater, seawater
8156BNUWP, 8157BNUMD, 8157BNURCA and 8302BNURCA
815600, 8165BNWP
9107APMD
9156APWP
9165BNWP
GD9156BNWP
9156DJWP
pH electrode catalog numbers by sample type cont.
14
Electrode
family
pH precision
Temperature range
Body type
ATC
Fill solution cat. no.
Special features
Cat. no.
ROSS ultra
0.01
0 to 100°C
Glass
810007
8102BNUWP
0.01
0 to 100°C
Glass
810007
Semi-micro
8103BNUWP
0.01
0 to 100°C
Glass
810007
Rugged bulb
8104BNUWP
0.01
0 to 60°C
Epoxy
Yes
N/A (Gel)
pH/ATC Triode
8107BNUMD, 8107BNURCA
0.01
0 to 100°C
Epoxy
810007
Semi-micro
8115BNUWP
0.01
0 to 100°C
Epoxy
810007
8156BNUWP
0.01
0 to 100°C
Epoxy
Yes
810007
pH/ATC Triode
8157BNUMD, 8157BNURCA
0.01
0 to 100°C
Glass
Yes
810007
pH/ATC Triode
8302BNUMD, 8302BNURCA
ROSS
0.01
0 to 100°C
Glass
810007
8102BN
0.01
0 to 100°C
Glass
810007
Semi-micro
8103BN
0.01
0 to 100°C
Glass
810007
Rugged bulb
8104BN
0.01
0 to 100°C
Epoxy
810007
Semi-micro
8115BN
0.01
0 to 100°C
Epoxy
810007
8135BN
0.01
0 to 100°C
Epoxy
810007
Flat surface
815600
0.01
0 to 100°C
Glass
810007
Spear tip
8163BNWP
ROSS Sure-Flow and Micro
0.01
0 to 100°C
Epoxy
810007
Sure-Flow
8165BNWP0
0.01
0 to 100°C
Glass
810007
Sure-Flow
8172BNWP
0.01
0 to 100°C
Epoxy
810007
Semi-micro
Sure-Flow
8175BNWP
0.01
0 to 100°C
Glass
810007
Micro,
3 x 40 mm tip
8220BNWP
No Cal
0.1
0 to 100°C
Epoxy
510011
Calibration stable for 1 year, pH/ATC
5107BNMD
AquaPro
0.02
0 to 60°C
Glass
910008-WA
KnipHe replacement
9121APWP
0.02
0 to 60°C
Glass
N/A (Polymer)
Rugged bulb
9104APWP
0.02
0 to 60°C
Epoxy
Yes
N/A (Polymer)
pH/ATC Triode
9107APMD
0.02
0 to 60°C
Epoxy
N/A (Polymer)
Flat surface
9135APWP
0.02
0 to 60°C
Epoxy
N/A (Polymer)
9156APWP
ROSS Ultra Triode
ROSS
ROSS Sure-Flow
ROSS Micro
Measurements at maximum temperature range should not be for extended periods of time.
pH electrode summary table
15
Thermo Scientific pH electrode summary table, cont.
Electrode
family
pH precision
Temperature range
Body Type
ATC
Fill solution cat. no.
Special features
Cat. no.
Standard Ag/AgCl
0.02
0 to 90 °C
Glass
900011
9102BNWP
0.02
0 to 90 °C
Glass
900011
Semi-micro
9103BNWP
0.02
0 to 90 °C
Glass
900011
Rugged bulb
9104BNWP
0.02
0 to 80 °C
Epoxy
Yes
Gel
pH/ATC Triode
9107BNMD
0.02
0 to 90 °C
Epoxy
900011
9156BNWP
0.02
0 to 90 °C
Epoxy
Yes
900011
pH/ATC Triode
9157BNMD
0.02
0 to 90 °C
Glass
900011
Low resistance
9157BNMD
0.02
0 to 90 °C
Glass
900011
Spear tip
9163SC
0.02
0 to 100 °C
Epoxy
900011
Sure-Flow
9165BNWP
0 to 100 °C
Glass
900011
Sure-Flow
9172BNWP
Green
0.05
0 to 90 °C
Epoxy
Gel
GS9106BNWP
0.02
0 to 90 °C
Epoxy
900011
GS9156BNWP
0.02
0 to 90 °C
Epoxy
Gel
Double junction
GS9106BNWP
0.02
0 to 90 °C
Epoxy
910008-WA
Double junction
GS9156BNWP
SJ Gen. Purpose
SJ Gel Triode
Sure Flow
DJ Semi-micro
Measurements at maximum temperature range should not be for extended periods of time.
16
pH electrode preparation
The following is a general procedure for preparing most pH electrodes. Refer to the electrode user guide or instruction manual for an electrode-specific preparation procedure.
1.
Remove the protective shipping cap, sleeve or bottle from the electrode pH-sensing bulb and save the cap, sleeve or bottle for storage. If the electrode has a storage bottle covering the pH-sensing bulb, unscrew the storage bottle cap before removing the electrode from the storage bottle.
2.
Clean any salt deposits from the electrode exterior by rinsing it with distilled water.
3.
If the electrode is refillable, uncover the fill hole and add the appropriate filling solution to the electrode. To maintain an adequate flow rate, the level of filling solution must always be above the reference junction and at least one inch above the sample level. The fill hole should be open whenever the electrode is in use.
4.
Gently shake the electrode downward (similar to a clinical thermometer) to remove any air bubbles that may be trapped inside the electrode.
5.
Soak the electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, or Orion pH Electrode Storage Solution, Cat. No. 910001, for at least 30 minutes. The ROSS Storage Solution is recommended for enhanced electrode performance.
6.
Connect the electrode to the meter.
7.
Select at least two pH buffers that bracket the expected sample pH and will be used during the calibration of the pH electrode.
pH electrode operation guide
Thermo Scientific pH Electrode Buffers and Storage Solutions
Description
Cat. No.
ROSS pH Electrode Storage Solution, 475 mL bottle
810001
ROSS All-in-One pH Buffer Kit, includes one 475 mL bottle each of 4.01, 7.00 and 10.01 buffer, one 475 mL bottle of ROSS pH electrode storage solution, one 60 mL bottle of pH electrode cleaning solution and one electrode storage bottle.
810199
ROSS All-in-One pH Buffer Kit, includes one 475 mL bottle each of 4.01, 7.00 and 10.01 buffer, one 475 mL bottle of standard pH electrode storage solution and one electrode storage bottle
910199
pH 1.68 buffer, 475 mL bottle
910168
pH 1.68 buffer, 5 x 60 mL bottles
9116860
pH 4.01 buffer, 475 mL bottle
910104
pH 4.01 buffer, 5 x 60 mL bottles
910460
pH 4.01 buffer, 25 x 15 mL individual buffer pouches
910425
pH 5.00 buffer, 475 mL
910105
pH 6.86 buffer, 475 mL bottle
910686
pH 6.86 buffer, 5 x 60 mL bottles
916860
pH 7.00 buffer, 475 mL bottle
910107
pH 7.00 buffer, 5 x 60 mL bottles
910760
pH 7.00 buffer, 25 x 15 mL individual buffer pouches
910725
pH 9.18 buffer, 475 mL bottle
910918
pH 9.18 buffer, 5 x 60 mL bottles
9191860
pH 10.01 buffer, 475 mL bottle
910110
pH 10.01 buffer, 5 x 60 mL bottles
911060
pH 10.01 buffer, 25 x 15 mL individual buffer pouches
911025-WA
pH 12.46 buffer, 475 mL bottle
910112
pH 12.46 buffer, 5 x 60 mL bottles
911260-WA
Standard pH electrode storage solution, 475 mL bottle
910001
Pure water pHISA low ionic strength adjuster, 5 x 60 mL
700003
Pure water pH 4.10 buffer B for low ionic strength samples, 4 x 475 mL
700402
Pure water pH 6.97 buffer A for low ionic strength samples, 4 x 475 mL
700702
Pure water pH 9.15 buffer C for low ionic strength samples, 4 x 475 mL
700902
pH 4.01 buffer, Twin-neck, 500 mL
BS04/500ML
pH 7.00 buffer, Twin-neck, 500 mL
BS07/500ML
pH 10.01 buffer, Twin-neck, 500 mL
BS10/500ML
17
Calibration recommendations
1.
Always pour fresh pH buffers into clean beakers for calibration. Choose buffers that are one to four pH units apart.
2.
Check the electrode slope daily by performing at least a two buffer calibration. The slope should be 92 to 102% (54.43 to 60.34 mV per pH unit).
3.
If the electrode is refillable, uncover the fill hole during calibration to ensure a uniform flow of filling solution. The filling solution level inside of the electrode must be at least one inch above the buffer solution level.
4.
The buffer solution level must be above the pH electrode reference junction when the electrode is immersed in the buffer.
5.
Between buffers, rinse the electrode with distilled water and then with the next buffer. To reduce the chance of error due to polarization, avoid rubbing or wiping the electrode bulb. Use a lint-free tissue and gently blot the bulb.
Use the Thermo Scientific™ Orion Star Stirrer Probe, Cat. No. 096019, or a magnetic stir plate and stir bar to stir all buffers and samples at a moderate, uniform rate. The stirrer probe can be used with the Orion Star A, Dual Star, and Versa Star Pro benchtop meters.
pH electrode calibration
pH Values of Buffers at Various Temperatures
Nominal
pH value
at 25 °C
0 °C
5 °C
10 °C
20 °C
30 °C
40 °C
50 °C
60 °C
70 °C
80 °C
90 °C
1.68
1.67
1.67
1.67
1.68
1.68
1.69
1.71
1.72
1.74
1.77
1.79
4.01
4.00
4.00
4.00
4.00
4.02
4.03
4.06
4.09
4.12
4.16
4.21
6.86
6.98
6.95
6.92
6.87
6.85
6.84
6.83
6.84
6.85
6.86
6.88
7.00
7.11
7.08
7.06
7.01
6.98
6.97
6.96
6.97
7.00
7.03
7.08
9.18
9.46
9.40
9.33
9.23
9.14
9.07
9.01
8.96
8.92
8.89
8.85
10.01
10.32
10.25
10.18
10.06
9.97
9.89
9.83
9.79
9.78
9.78
9.80
12.46
13.47
13.24
13.03
12.64
12.29
11.99
11.73
11.50
11.30
11.13
10.98
1.
If using a magnetic stir plate, place a piece of insulating material, such as polystyrene foam or cardboard, between the magnetic stir plate and beaker to prevent measurement errors from the transfer of heat to the sample.
2.
Handle the micro pH electrodes with care. Do not touch the pH bulb and stem against the bottom or walls of the sample containers.
Three buffer calibration procedures
This procedure is recommended for precise measurements. Refer to the meter user guide for detailed instructions on how to perform a pH calibration using your meter.
1.
Verify that the pH electrode was prepared correctly and connect the electrode to the meter.
2.
Select three pH buffers that bracket the expected sample pH. It is recommended that one buffer be near the electrode isopotential point (pH 7) and the other buffers be near the expected sample pH and one to four pH units apart (for example, pH 4 and pH 10). The pH buffers should be at the same temperature as the sample. If the buffers and samples are at varying temperatures, temperature compensation is recommended.
18
8.
When the reading is stable, accept the buffer value using the meter’s automatic buffer recognition feature or manually enter the value of the pH buffer at its measured temperature. Rinse the electrode with de-ionized water and gently blot excess drops with
a lint-free tissue.
9.
Place the electrode into the third buffer, so the electrode tip and junction are fully immersed in the buffer, and stir the buffer at a moderate, uniform rate.
10.
When the reading is stable, accept the buffer value using the meter’s automatic buffer recognition feature or manually enter the value of the pH buffer at its measured temperature.
11.
The meter should display a 92 to 102% slope or 54 to 60 mV per pH unit, depending on the pH meter. Refer to the meter user guide for details on how the meter displays the calibration information.
1.
Pour about 30 mL of each pH buffer into 50 mL beakers and proceed with the calibration immediately.
2.
Start the calibration on the meter.
3.
Rinse the electrode with de-ionized water and gently blot excess drops with a lint-free tissue.
4.
Place the electrode into the first buffer, so the electrode tip and junction are fully immersed in the buffer, and stir the buffer at a moderate, uniform rate.
5.
When the reading is stable, accept the buffer value using the meter’s automatic buffer recognition feature or manually enter the value of the pH buffer at its measured temperature. The table on page 16 provides pH values for Orion Buffers at various temperatures.
6.
Rinse the electrode with de-ionized water and gently blot excess drops with a lint-free tissue.
7.
Place the electrode into the second buffer, so the electrode tip and junction are fully immersed in the buffer, and stir the buffer at a moderate, uniform rate.
pH electrode calibration (continued)
19
General pH measurement recommendations
•
Check the electrode slope daily by performing at least a two buffer calibration. The slope should be 92 to 102%.
•
Unless otherwise specified, only use the recommended filling solution in refillable pH electrodes. The pH Electrode Summary Table section shows the electrode filling solutions that are recommended for the Thermo Scientific pH Electrodes.
•
If the electrode is refillable, uncover the fill hole during measurements to ensure a uniform flow of filling solution. The filling solution level inside of the electrode must be at least one inch above the sample solution level.
•
The sample solution level must be above the pH electrode reference junction when the electrode is immersed in the sample.
•
Between measurements, rinse the electrode with distilled water and then with the next solution to be measured. To reduce the chance of error due to polarization, avoid rubbing or wiping the electrode bulb. Use a lint-free tissue and gently blot the bulb.
•
Use the Orion Star Series Stirrer Probe, Cat. No. 096019 or a Magnetic Stir Plate and Stir Bar to stir all buffers and samples at a moderate, uniform rate. The stirrer probe can be used with Orion Star A, Dual Star and Versa Star Pro benchtop meters.
•
If using a magnetic stirrer, place a piece of insulating material, such as polystyrene foam or cardboard, between the magnetic stir plate and beaker to prevent measurement errors from the transfer of heat to the sample.
•
If the electrode is refillable and the electrode is used in dirty or viscous samples or the electrode response becomes sluggish, empty the electrode completely and hold the reference junction under warm running water. Empty any water from the electrode and fill it with fresh filling solution. For a more thorough cleaning, refer to the pH Electrode Cleaning Procedures section.
•
Flat surface electrodes may be used on any moist surface or in liquids.
•
Handle the micro pH electrodes with care. Do not touch the pH bulb and stem against the bottom or walls of the sample containers.
pH electrode measurements
USP <791> measurement requirements & recommendations
Orion™ pH meters, electrodes and solutions are part of a high-quality pH test method designed to assist with compliance to USP <791> pH requirements. Orion pH meters, electrodes and buffers offer features commonly desired by highly-regulated labs.
USP <791>
General Recommendation*
Temperature Device: ±1 ºC
Verify against a NIST-traceable thermometer
Calibration Buffers: 2 Minimum
(USA Buffer Set)
3-point calibration with pH 4.01, 7.00, and 10.01 buffers, or
5-point calibration with pH 1.68, 4.01, 7.00, 10.01, and 12.46 buffers
Slope Guideline Value: 90 - 105 %
Slope is displayed on the meter screen when calibration is complete
Slope is also displayed in the meter calibration log
Offset Guideline: ±30 mV
Offset can be calculated by the meter, or
Offset can be measured in pH 7 buffer, or
Offset can be calculated by the analyst
Verification Requirement: ±0.05 pH
Analyze pH 6.86 buffer
Compare results to the table of buffer values at temperature at the bottom of page 5
* Consult with the SOP prepared by your internal quality or regulatory group when performing your testing per USP <791>.
Slope Value of 92-102% and Offset of ±59 mV are defined by Thermo Scientific Orion system specifications.
20
Sample requirements
Electrodes with an epoxy body should only be used
in aqueous solutions.
Electrodes with a glass body may be used in nonaqueous solutions and solutions that contain organic solvents. A minimum of 20% water must be present in the sample for the best measurement results.
A benefit of the ROSS Ultra and ROSS Electrodes is that the filling solution composition may be changed depending on the sample requirements. The ROSS Electrode Filling Solution, Cat. No. 810007, is 3 M KCl. If there is a great deal of drift when using a ROSS Ultra or ROSS Electrode in nonaqueous solutions or solutions that contain organic solvents, fill the electrode with a mixture of methanol and water saturated with KCl. The ratio of methanol to water should be similar to the sample composition. For solutions that precipitate in the presence of chloride ions, fill the ROSS Ultra or ROSS Electrode with 10% KNO3, Cat. No. 900003.
Standard Ag/AgCl electrodes, micro Ag/AgCl electrodes and economy electrodes contain a single junction silver/silver chloride reference that will become clogged in solutions that contain silver complexing or binding agents such as TRIS buffer, proteins and sulfides. Frequent cleaning may be required when measuring these solutions, which will shorten the electrode life. To better measure pH in these solutions, use the ROSS Ultra electrodes, ROSS Electrodes, AquaPro Electrodes or Double Junction Electrodes. Proteins cause the additional problem of coating the pH-sensing bulb, so extra care should be taken to keep the electrode clean while measuring samples that contain proteins. Regular use of the Orion pH Electrode Cleaning Solution A for removal of proteins is recommended. See page 15 for details.
pH measurement procedure
1.
Verify that the pH electrode was prepared and calibrated correctly. If the electrode is refillable, make sure that the fill hole is uncovered and the filling solution level is at least one inch above the sample solution level.
2.
Rinse the electrode with distilled water and blot it dry with a lint-free tissue.
3.
Place the electrode into the sample. Use the Thermo Scientific stirrer probe or a magnetic stir plate
and clean stir bar to stir the sample at a moderate, uniform rate.
4.
When the measurement is stable, record the pH value and temperature of the sample.
21
1.
On a weekly basis, inspect the pH electrode for scratches, cracks, salt crystal build-up, or membrane/junction deposits.
2.
Rinse off any salt build-up with distilled water. Remove any membrane/junction deposits by soaking the electrode in a 0.1 M HCl or 0.1 M HNO3 solution for 15 minutes or soaking the electrode in a 0.1 M KCl solution heated to 55 °C for 15 minutes. If the electrode is dirty, clogged or coated, refer to the pH Electrode Cleaning Procedures section for a more thorough electrode cleaning procedure.
3.
If a refillable electrode is used, drain the reference chamber, flush it with distilled water until all of the salt crystal build-up inside of the electrode is removed, flush it with fresh filling solution and fill the reference chamber with fresh filling solution.
4.
Soak the electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, or Orion pH Electrode Storage Solution, Cat. No. 910001, for 1 to 2 hours. The ROSS pH Electrode Storage Solution is recommended for enhanced electrode performance.
pH electrode maintenance
Filling and draining a refillable pH electrode
To fill an electrode, install the flip spout cap onto the filling solution bottle and lift the flip spout to a vertical position. Insert the spout into the electrode fill hole and add filling solution up to the fill hole. If a Sure-Flow electrode is used, push down on the electrode cap to allow a few drops of filling solution to drain out of the electrode and release the cap to reset the Sure-Flow junction. Push down and release the cap until the junction returns to its original position and add filling solution up to the fill hole.
To drain most electrodes, insert a lint-free tissue or plastic pipet into the fill hole and remove all of the filling solution. If a pipet is used, do not insert the pipet too far into the electrode and do not touch the inner glass tube with the pipet.
To drain a Sure-Flow pH electrode, make sure that the fill hole is uncovered, place a waste beaker under the electrode and push down on the electrode cap to remove all of the filling solution. The filling solution will flow out of the Sure-Flow junction, near the pH bulb.
22
To ensure a quick electrode response and unclogged electrode junction, the electrode should never be stored dry and the pH-sensing bulb and reference junction must not dry out. Always store the pH electrode in pH electrode storage solution.
Short term electrode storage (up to one week)
Soak the electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, or Orion pH Electrode Storage Solution, Cat. No. 910001. For enhanced electrode performance, use ROSS pH Electrode Storage Solution, The Thermo Scientific™ Swing Arm Electrode Stand, Cat. No. 090043, is pictured on the right.
To minimize the chance of breakage, the micro pH electrodes should be attached to an electrode stand and suspended in a beaker that contains storage solution. The electrode should not touch the sides or bottom of the beaker.
Long term electrode storage (more than one week)
If the electrode is refillable, fill the reference chamber up to the fill hole with the appropriate electrode filling solution and securely cover the fill hole with parafilm. Cover the pH-sensing bulb and reference junction with a protective cap, sleeve or storage bottle that contains storage solution. Before returning the electrode to use, prepare it as a new electrode.
pH electrode storage
Thermo Scientific pH Electrode Storage Solutions and Accessories
Description
Cat. No.
Standard pH Electrode Storage Solution, 475 mL bottle
910001
Standard pH Electrode Storage Solution, 5 x 60 mL bottles
910060
ROSS pH Electrode Storage Solution, 475 mL bottle
810001
Storage Bottles for electrodes with a 12 mm diameter, pack of 3
910003
Storage Bottles for electrodes with an 8 mm diameter, pack of 3 (semi-micro epoxy pH electrodes)
910004-WA
Storage Bottles for electrodes with a 6 mm diameter, pack of 3 (semi-micro glass pH electrodes)
910006-WA
Swing Arm Electrode Stand
090043
The Orion Storage Sleeve and Base is for use with pH electrodes that are 120mm or less in length that have a diameter of 12mm. The sleeve protects your electrode from breakage when not in use and it won’t tip over due to its weighted removable base. As an additional convenience, the pH bulb stays conditioned while in storage so that it’s ready to use when removed from the sleeve.
810017
23
General troubleshooting procedures
Follow a systematic procedure to isolate the problem. The pH measuring system can be divided into four components: pH meter, electrode, sample/application
and technique.
pH meter
The meter is the easiest component to eliminate as a possible cause of error. Thermo Scientific pH Meters include an instrument checkout procedure and BNC shorting cap for convenience in troubleshooting. Refer to the pH meter user guide for details.
Electrode
To test the electrode operation:
1.
Connect the electrode to a working meter that has a mV measuring mode. Set the meter to the mV mode.
2.
Rinse the electrode with distilled water, blot it dry with a lint-free tissue and insert the electrode into fresh pH 7 buffer. When the reading is stable, record the mV value of the pH 7 buffer. The mV value should be -30 to +30 mV.
3.
Rinse the electrode with distilled water, blot it dry with a lint-free tissue and insert the electrode into fresh pH 4 buffer. When the reading is stable, record the mV value of the pH 4 buffer. The mV value should be +150 to +210 mV.
4.
Calculate the absolute mV difference between the two buffers. The absolute mV difference should be 160 to 180 mV. The actual mV values will change as the electrode ages, but the mV difference between the two buffers should always be 160 to 180 mV.
If the electrode fails this procedure, open new pH buffers and repeat the procedure, clean the electrode thoroughly as directed in the “pH Electrode Cleaning Procedures section” and soak the pH electrode overnight in storage solution. If the electrode is refillable and the electrode response is slow or drifting, drain and refill the electrode with fresh filling solution. Replace the electrode if cleaning and maintenance fail to rejuvenate it.
Sample/application
The electrode may work properly with pH buffers, but not with the sample. In this case, check the sample composition for interferences, incompatibilities or temperature effects.
Technique
If trouble persists, review operating procedures. Review the pH Electrode Calibration section and pH Electrode Measurements section to be sure proper technique has been followed.
pH electrode interferences
Sodium ions are the principal interference of the pH electrode and cause increasing error at higher pH (lower hydrogen ion activities) and at higher temperatures. Because the pH membrane is composed of special low sodium error glass, error due to sodium is negligible when measuring at pH values less than 12. When measuring at pH values greater than 12, add the correction value from the nomograph below to the observed pH reading.
pH electrode troubleshooting guide
Sodium Error Example
pH reading 12.10
Sodium concentration 0.5
MTemperature 50 °C
Correction 0.01
Corrected pH reading 12.11
10
111213807060504030°C 20 0.1 M250.5 M1 M2 M5 M5 M0.5 M0.1 M1 M2 M20°C253040506070800.700.600.500.400.300.250.200.150.100.080.060.050.040.030.020.010.005pH correction to be addedSamplepH readingsamplesodiumconcentration
24
One of the most common reasons for a pH electrode to not work properly is because it is dirty, clogged or coated with sample. Cleaning a dirty, clogged or coated electrode restores proper electrode performance and prolongs the useful life of the electrode.
The Thermo Scientific pH Electrode Cleaning Solutions are designed to simplify pH electrode maintenance and the ready-to-use cleaning solutions include a small beaker to hold the cleaning solution and a plastic pipet for removing the electrode filling solution. The Orion pH Electrode Cleaning Solution Kit, Cat. No. 900020, features one bottle of each cleaning solution for operators who are working with a variety of sample matrices.
Optimal cleaning procedures are dependent upon the sample type, extent of build-up or clogging and the type of electrode. The following instructions provide a starting point for developing an effective cleaning procedure.
1.
Choose a cleaning solution. Cleaning solution D is a mild cleaning solution and cleaning solution C is a stronger cleaning solution. Cleaning solution A is for removing protein deposits and cleaning solution B is for removing bacterial contaminants.
2.
Shake the cleaning solution. Pour enough of the cleaning solution into the beaker to cover the electrode junction. Refer to Figure 1 on page 5 for the location of typical pH electrode components, such as the electrode junction.
3.
Soak the electrode for 5 to 10 minutes in the cleaning solution while moderately stirring the solution. Electrodes with wick junctions may require more cleaning time.
4.
Remove the electrode from the cleaning solution and rinse it thoroughly with distilled water to remove all traces of the cleaning solution.
5.
If cleaning a refillable electrode, remove the filling solution from the electrode using the pipet that is included with the kit and add fresh filling solution to the electrode. Repeat this step two or three times for optimal electrode performance.
6.
If cleaning a Sure-Flow electrode, flush a few drops of filling solution through the electrode junction by pressing down on the electrode cap. Ensure that the junction flushes and resets properly. Refill the electrode with filling solution.
7.
Soak the electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, or Orion pH Electrode Storage Solution, Cat. No. 910001, for at least 30 minutes.
8.
Rinse the electrode thoroughly with distilled water and measure samples as usual. If the electrode response is slow or the electrode does not calibrate correctly, repeat the cleaning procedure. Viscous samples and samples that contain solid materials often require additional cleaning and additional filling solution changes.
pH electrode
cleaning procedures
Thermo Scientific pH Electrode Cleaning Solutions
Description
Cat. No.
pH electrode cleaning solution kit, 1 x 30 mL bottle each of cleaning solution A and C, 1 x 60 mL bottle each of cleaning solution B and D, beaker and pipet
900020
pH electrode cleaning solution A, removes protein contaminants, 4 x 30 mL bottles, beaker and pipet
900021-WA
pH electrode cleaning solution B, removes bacterial contaminants, 4 x 60 mL bottles, beaker and pipet
900022-WA
pH electrode cleaning solution C, for general cleaning, 4 x 30 mL bottles, beaker and pipet
900023
pH electrode cleaning solution D, removes oil and grease contaminants, 4 x 60 mL bottles, beaker and pipet
900024
25
Using high-quality solutions to calibrate and maintain your sensors is the best way to ensure your data is both accurate and reproducible. Avoid using low-quality, home-made, or expired buffers. These can lead to measurement errors, which require lengthy troubleshooting, or may go unnoticed. When choosing or making your own buffers and solutions for pH measurement, pay attention to the quality of the water and chemicals used. Use USP grade water whenever possible. Orion Solutions are made with USP-grade purified water and are traceable to NIST certified standards. They undergo rigorous quality testing throughout the process to prevent contamination before and after bottling.
Role of solution quality in pH measurement
Unopened Thermo Scientific pH Buffers and Electrode Filling Solutions are NIST traceable. All buffers, except for pH 10.01, have an expiration date of two years from the date of manufacturing. pH 10.1 buffer has an expiration date of 18 months from date of manufacture. Certificates of analysis, which include the expiration date, are available at thermofisher.com.
Once a pH buffer or electrode filling solution bottle is opened, discard the unused portion after 2 to 3 months. Once pH 10.01 buffer is opened, discard the unused portion after 1 month, since pH 10.01 buffer is vulnerable to carbon dioxide contamination. Never pour used pH buffers or electrode filling solutions back into the bottle. Should buffers inadvertently become frozen during storage, allow buffers to come to room temperature before use.
Recommended shelf life for buffers and solutions
26
pH measurements in TRIS buffer, protein and sulfide samples
TRIS [tris (hydroxymethyl) aminomethane] is a compound frequently used as a biological buffer, sulfide is common in wastewater and petroleum products and proteins are often in food products, wastewater and biological samples. When a single junction Ag/AgCl pH electrode is used in TRIS buffer, protein or sulfide samples, the sample will complex with the silver in the electrode and clog the electrode junction, preventing stable pH measurements. Proteins also coat the pH-sensing bulb. For the most accurate results and most stable measurements in these samples, use a pH electrode with a ROSS or double junction Ag/AgCl reference. The ROSS Ultra, ROSS, AquaPro and Double Junction pH Electrodes are recommended for measuring the pH in TRIS buffer, protein or sulfide samples. When measuring protein samples, soak the pH electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, between measurements to remove the protein coating. When cleaning the electrode is required, use Orion pH Electrode Cleaning Solution A, Cat No. 900021-WA, for protein removal.
pH measurements in solid samples and surface pH measurements
Solid and semi-solid samples include cheese, meat, powders, paper and agar gels. There are many methods available for measuring the pH of solid and semi-solid samples that include using a flat surface pH electrode, using a spear tip pH electrode, and mixing or blending a fixed amount of sample with distilled water. Once a method is established that is compatible with the sample, the method should be followed exactly with all samples for the most reproducible measurements.
If the sample is soft or semi-solid, use a spear tip pH electrode and pierce the sample with the electrode to a fixed depth. If the surface of the sample is moist, use a flat surface pH electrode and measure the surface of the sample. The sample must be moist enough for the pH-sensing bulb and the reference junction to make adequate contact with the sample. If necessary, add a drop of distilled water or potassium chloride to wet the surface
Common pH sample issues and remediesbefore placing the electrode on the sample. For the best reproducibility, all samples should have the same amount of liquid added before measurement.
If the sample is soluble in water or disperses readily in water, mix a fixed percent of sample with a fixed quantity of water and measure the resulting solution. Up to 25% sample by weight or volume is commonly used. Since each sample is different, the details of the method should be determined through trial and error with the sample. For cheese samples – blend 5 grams of cheese with 100 mL of distilled water and stir the solution thoroughly, stop stirring and wait 5 minutes to let the solids settle and then measure the pH of the liquid portion. For paper samples – soak 1 gram of paper pieces in distilled water, wait 10 minutes and measure the pH of the liquid. For soil samples – mix a fixed ratio of soil and distilled water and stir the solution thoroughly, stop stirring and wait a fixed amount of time to let the solids settle and then immerse the pH electrode at a fixed depth and measure the pH of the liquid portion.
pH measurements in pure water samples
Pure water samples include distilled water, deionized water, some process waters, well water, some surface waters, treated effluent, boiler feed water and rain water. In pure water samples the pH electrode response tends to drift and is slow, non-reproducible and inaccurate. Measurements in these samples are more difficult because of the low conductivity of the sample, differences between the low ionic strength solutions and normal ionic strength buffers, changes in the liquid junction potential and the absorption of carbon dioxide into the sample. Since pure water solutions have a low conductivity, the solution will tend to act as an antenna and the electrode response can be noisy.
Calibration of a pH electrode in high ionic strength buffers will increase the time required for stabilization when measuring a low ionic strength sample. In addition, the possibility of sample contamination will be increased. For most precise measurements, buffers and samples should have a similar ionic strength. When two solutions come in contact, diffusion occurs until an equilibrium is reached.
27
Since ions have different mobility and diffuse at different rates, a charge imbalance occurs at the point of contact. A liquid junction potential occurs when the electrode filling solution meets the sample. This charge imbalance will be large if there is a large difference in composition between the filling solution and the sample. It is important that the junction potential be constant during measurement. If the two solutions are quite different, normal fluctuations in the boundary layer will produce noise. Constant, reproducible junction potentials are achieved by measuring samples and standards with similar ionic strengths. Since pure waters contain little dissolved material, their buffering capacity is small. Absorption of carbon dioxide from the atmosphere will cause a slow change in pH, observed as a drifting pH reading and a different pH from the original sample. Samples that are not previously saturated with carbon dioxide must be handled with care.
The Thermo Scientific™ Pure Water™ pH Test Kit, Cat. No. 700001, has been developed to minimize the problems encountered when measuring pH in pure waters. This test kit uses a pure water pH additive called a pHISA™ adjustor and low ionic strength buffers that contain the same background as the pHISA adjustor. For the most accurate results, a ROSS Ultra or ROSS pH Electrode is recommended. Adding pH ISA adjustor to samples increases the ionic strength, reducing the noise and improving the response time. The shift in pH caused by the addition of pHISA adjustor is minimal – between 0.005 and 0.01 pH units. Since the same amount of pHISA adjustor is added to the buffers and samples, the net effect on the pH is negligible. Calibration is performed using the low ionic strength buffers with a pH ISA adjustor already added. Measuring with samples and buffers with the same ionic strength increases accuracy, precision and time response. Contamination due to carryover from higher ionic strength buffers is also minimized.
pH measurements in sludges, suspensions,
colloids, slurries and viscous samples
Sludge, suspension, colloid, slurry and viscous samples include wastewater, mud, paper pulp and corn syrup. These samples clog the electrode junction and coat the pH-sensing bulb, resulting in slow electrode response, measurement drift and pH measurement errors. The thickness of the sample can also cause a sample carryover problem – after the pH electrode is removed from the sample, some of the sample may cling to the electrode.
For the most accurate results and most stable measurements in these samples, use a pH electrode with a Sure-Flow or open junction and soak the pH electrode in ROSS pH Electrode Storage Solution, Cat. No. 810001, between measurements. For the best reproducibility, immerse the pH electrode at a fixed depth in all samples. Sure-Flow junctions resist clogging, provide a constant flow of filling solution and are extremely easy to clean. Open junctions resist clogging, contain a low maintenance polymer gel as the filling solution and are easy to clean. The ROSS Sure-Flow, Ag/AgCl Sure-Flow and AquaPro pH Electrodes are recommended for measuring the pH in sludge, suspension, colloid, slurry and viscous samples. The ROSS pH Electrode Storage Solution will enhance the pH electrode performance by acting as a very mild cleaning solution. For a more thorough cleaning procedure, refer to the “pH Electrode Cleaning Procedures section”.
pH measurements in extreme
pH or high salt content samples
Samples with a pH less than 2, pH greater than 12 and salt content greater than 0.1 M include battery acid, copper plating solutions and brines. These samples create measurement issues with the reference portion of the pH electrode, due to the liquid junction potential. The liquid junction potential occurs when there is a large difference in composition between the filling solution and the sample, resulting in measurement drift and slow electrode response.
In samples with a pH of 2 to 12 or salt content less than 0.1 M, the liquid junction potential is rarely a problem since the filling solution is formulated for the best performance under these conditions. In samples with a pH less than 2, pH greater than 12 or a salt content greater than 0.1 M, a pH electrode with a ROSS or double junction Ag/AgCl reference is recommended. The filling solution of the ROSS Ultra, ROSS and Refillable Double Junction pH Electrodes can be modified to match the sample composition, which will reduce the liquid junction potential and provide accurate, more stable measurements in these samples.
For samples with a pH greater than 12, add a dilute base to the recommended electrode filling solution to raise the filling solution pH. For samples with a pH less than 2, add a dilute acid to the recommended electrode filling solution to lower the filling solution pH. For samples with a high salt content, use a strong solution of the same salt as the electrode filling solution. For example, when measuring the pH of sodium bromide brines, use sodium bromide in the outer chamber of a double junction electrode.
Find out more at thermofisher.com/pH
This product is intended for General Laboratory Use. It is the customer’s responsibility to ensure that the performance of the product is suitable for customers’ specific use or application. © 2014-2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. S-PHREFBK-E 0720 RevD
pH measurements in nonaqueous samples
Nonaqueous samples contain liquids other than water and include oils, alcohols, cosmetics, methanol and acetone. These samples produce unstable readings, slow electrode response and pH measurement errors due to high sample resistance, bulb dehydration, sample carryover contamination and liquid junction potentials. The high resistance (low conductivity) of many organic solvents causes noise and slow electrode response. To avoid this, an electrode with a low resistance glass bulb should be used. It is often helpful to add a small amount of an inert salt (for example, a quaternary ammonium salt) to increase the conductance of the solution. Addition of a salt will change the activity of the hydrogen ion, causing a small shift in the pH; however, the error is small when compared with measurement drift issues.
Slow electrode response and drift are also caused by the dehydration of the glass pH sensing bulb. For the pH sensing bulb to function, the surface of the glass must be hydrated so hydrogen ions can be absorbed onto the surface of the glass. A nonaqueous solution causes the dehydration of the pH sensing bulb and frequently soaking the pH electrode in water or a pH buffer hydrates the bulb, providing a faster, more stable response. To prevent contamination of samples due to carryover on the electrode bulb, the electrode should be rinsed between measurements with a solvent that will dissolve the sample material from the electrode, and then rinsed with a volatile solvent like acetone to remove the rinse solution, followed by soaking it in a pH buffer.
Large liquid junction potentials that cause unstable readings are a problem when measuring nonaqueous samples. When a reference electrode with an aqueous filling solution comes in contact with a nonaqueous sample, unequal diffusion of ions occur, causing continuous measurement drift. To minimize this problem, the electrode filling solution should be modified to provide compatibility with the sample and minimize junction potential problems. A refillable electrode with a double junction reference, such as a ROSS Ultra or ROSS pH Electrode, should be used to allow the filling solution to be easily modified. Several alternative filling solutions can be used and include a mixture of methanol and water saturated with KCI, glacial acetic acid saturated with Kl, acetone saturated with KI and methanol saturated with LiCI. When using a solvent in the filling solution, a salt must be added to improve the electrical conductance. KI is often more soluble than KCI and can be substituted as required. When adjusting the filling solution, allow the electrode to reach equilibrium overnight before using it. Solvent-based filling solutions should not be used in epoxy body electrodes. Since epoxy body electrodes should not be used in certain solvents, especially highly polar organic solvents, a glass pH electrode should be used.
The pH value of a nonaqueous solution cannot be accurately compared to an aqueous buffer. The activity of the hydrogen ion varies, depending on the background medium, due to the differences in dielectric constants, solvent acidities and ion mobilities. The pH values obtained when measuring nonaqueous samples can only be used as relative measurements, to compare the acid-base qualities of similar solvents or to indicate when adjustments in pH are needed. The readings cannot be reported as actual pH values, since the pH scale is based on the relative acidity or alkalinity of water.