In wastewater analysis, total organic carbon and total bound nitrogen (TOC/TNb ) can be determined with a cellulose test. For this, TOC/TNb analyzers must ensure the complete transfer and oxidation of the sample.
However, municipal wastewater, industrial effluents and surface-water have one challenge in common: particles which might lead to clogging of sample paths or sample carryover.
This application note highlights an advanced TOC/TNb analyzer that enables fast, reliable and cost-efficient measurement in highly particle loaded samples while meeting DIN EN ISO 20236 and DIN EN 1484 regulatory requirements.
Download this application note to learn:
- The challenges of particle-rich sample testing
- How to ensure accurate measurement of both dissolved and particle-bound fractions
- How robust particle handling ensures efficient analysis and regulatory compliance
Application Note · multi N/C 2300 & multi N/C 3300
Challenge
TOC determination in cellulose suspension with undissolved particles at a size of up to 100 µm according to DIN EN 1484 or DIN EN ISO 20236
Solution
Fast, reliable, and cost-efficient measurement in highly particle loaded samples thanks to the excellent particle handling of the combustion-based analyzers of the multi N/C x300 series.
TOC Recovery in Particle-Containing Samples with the Cellulose Test
According to DIN EN ISO 20236 and DIN EN 1484
Introduction
When monitoring total organic carbon (TOC) and total
bound nitrogen (TNb
) in environmental samples like surface
or waste water, by definition not only the dissolved organic
carbon and nitrogen compounds have to be measured, but
also the particle-bound fraction of these compounds have to
be fully detected. According to DIN EN 1484[1] and DIN EN
ISO 20236[2], the TOC is defined as the sum of all organically
bound carbon present in water, both in dissolved and
suspended matter. On the other hand, the dissolved organic
carbon (DOC) is defined as the sum of the organically
bound carbon in water originating from compounds
passing through a membrane filter of 0.45 μm pore size.
Accordingly, a TOC/TNb
analyzer used for the determination
of particulate environmental samples has to prove that both
sample homogenization and sample handling in the analyzer
from sample vial up to the combustion process ensures
complete transfer and oxidation of a representative aliquot
of the particulate sample.
To verify this, Annex C of DIN EN 1484 and chapter 8.2
of DIN EN ISO 20236 include a specific test to determine
the recovery and variation of replicate determination for
particle processing of a cellulose suspension with up to
100 µm particle size at a 100 mg/L TOC concentration. The
mean value from a triplicate measurement must not exceed
±10% of the theoretical value. The repeatability variation
coefficient must be ≤ 10%.
In addition to the necessary sample homogenization and
sample processing preconditions of a TOC/TNb
system, the
oxidation power of the analyzer is also critical to pass this
so-called cellulose test. At this point it has to be mentioned
that conventional wet-chemical and/or UV-based TOC
analyzers show only low recovery rates on this cellulose test
and are therefore not suitable for the TOC measurement of
typical particle-containing environmental water samples.
2 TOC Recovery in Particle-Containing Samples with the Cellulose Test According to DIN EN ISO 20236 and DIN EN 1484
Materials and Methods
Samples and reagents
To prepare a 100 mg/L C suspension solution, 225 mg
cellulose (particle size between 20 µm and 100 µm) is
placed in a 1000 mL volumetric flask and made up to the
mark with water. For proper homogenization, the solution
should then be stirred on a magnetic stirring plate, possibly
overnight to allow the micro cellulose to swell. Before
each use, the suspension must be stirred again in order
to prepare for a homogeneous solution. The filling of the
respective sample vessel plays a very important role. It is
absolutely necessary that the suspension solution is shaken
vigorously immediately before filling and then transferred
into the vessel relatively quickly with a large volume surge
until the vessel overflows. Alternatively, a large 5 or 10 mL
micropipette can be used to transfer the thoroughly mixed
suspension into sampler vials.
■ Wastewater samples were taken from the inlet as well as
from an intermediate treatment stage of a sewage plant
■ 2 M HCl was used for automatic sample acidification to a
pH value < 2
Sample preparation and measurement
The samples were stored in a refrigerator at 4 °C until
analysis and then transferred into suitable autosampler vials
after appropriate stirring. The wastewater samples and the
cellulose test suspensions were analyzed in direct mode
applying an NPOC method. The samples were adjusted
to pH < 2 using 2 M HCl, and subsequently purged for a
period of 5 minutes. Using an autosampler, representative
sample aliquots were transferred into the combustion
tube. The samples were catalytically oxidized at a furnace
temperature of 800 °C in an oxygen-rich atmosphere. A
16 mm combustion tube filled with platinum catalyst was
used. The CO2 formed was quantitatively detected by the
focus radiation non-dispersive infrared detector (FR-NDIR).
Optionally, the formed nitrogen oxides can be detected
by means of a chemiluminescence detector (CLD) or an
electrochemical detector (ChD).
Calibration
For the determination of the cellulose suspensions and
the samples, the TOC/TNb
analyzer was calibrated with
potassium hydrogen phthalate. Standards with the following
concentrations were prepared and measured: 50 mg/L,
100 mg/L, 250 mg/L, and 500 mg/L. Each calibration
point was injected with three replicates as agreed in the
method. The method characteristics resulting from the linear
regression are shown in the following diagram.
Figure 1: Calibration curve and method characteristics
3 TOC Recovery in Particle-Containing Samples with the Cellulose Test According to DIN EN ISO 20236 and DIN EN 1484
Instrumentation
The TOC measurements were performed on multi N/C 2300
and multi N/C 3300. Following method settings were used
to determine the TOC content.
The proper alignment of the sample aspiration needle on
the autosampler is important for a representative sample
aspiration of particle-containing samples, as magnetically
stirred water sample forms a stirring funnel. Sample take-up
is optimized when the sample aspiration needle is located
outside the center of the vial and, depending on the vial size,
at least 5 to 7 mm above the magnetic stirring bar. It is also
important to set the correct stirring speed in the method
in order to achieve good homogenization on the one hand,
but not to create a large stirring funnel at the same time.
Therefore, depending on the sample vessel size, the settings
recommended in the table above should be used.
Results and Discussion
The analysis results of all samples and cellulose test suspensions are summarized in Table 2. Measurements were performed
as triplicates. The achieved measurement reproducibility was within the expected range. A typical measuring curve of a
cellulose suspension with the corresponding replicate results is shown in Figure 2.
multi N/C 2300 multi N/C 3300
Parameter NPOC (direct TOC measurement) NPOC (direct TOC measurement)
Digestion High temperature combustion with Pt catalyst at 800 °C High temperature combustion with Pt catalyst at 800 °C
Number of repetitions Min. 3, max. 3 Min. 3, max. 3
Autosampler and vial sizes AS 60, 8 mL vials, 250 µL syringe AS vario ER, 72 pos. rack, 40 mL vials
Sample rinses before injection 3 3
Reverse rinses - 1
Sample injection volume 250 µL 500 µL
Stirring speed 5 7
NPOC purge time 300 s 300 s
Table 1: Method settings
Table 2: Results
Sample ID Result: NPOC [mg/L] RSD [%]
Sewage plant inflow 284.7 0.86
1st Cellulose test at 100 mg/L 94.9 0.41
Intermediate treatment 125.2 1.44
2nd Cellulose test at 100 mg/L 99.3 0.48
KHP check standard at 150 mg/L 150.3 0.10
This document is true and correct at the time of publication; the information within is subject to change. Other documents may supersede this document,
including technical modifications and corrections.
Trademark notice: The brand names of the third-party products specified in the application note are usually registered trademarks of the respective companies or organizations.
4 TOC Recovery in Particle-Containing Samples with the Cellulose Test According to DIN EN ISO 20236 and DIN EN 1484
Headquarters
Analytik Jena GmbH+Co. KG
Konrad-Zuse-Strasse 1
07745 Jena · Germany
Phone +49 3641 77 70
Fax +49 3641 77 9279
info@analytik-jena.com
www.analytik-jena.com
Version 2.0 · Author: BeBl
en · 09/2024
© Analytik Jena GmbH+Co. KG | Pictures ©: pixabay/JoeBreuer (p. 1)
References
[1] DIN EN 1484 Water analysis – Guidelines for the determination of total organic carbon (TOC) and dissolved organic carbon (DOC)
[2] DIN EN ISO 20236: Water quality — Determination of total organic carbon (TOC), dissolved organic carbon (DOC) total bound nitrogen (TNb
) and dissolved
bound nitrogen (DNb
) after high temperature catalytic oxidative combustion
Summary
The measurement results of the cellulose suspensions
exceed the required expectations of a recovery in the range
from 90 to 110% with results of 95% and 99%. Likewise,
the coefficients of variation at 0.41% and 0.48% exceed
the requirements of DIN EN 1484 or DIN EN ISO 20236,
respectively. Decisive for these good results are the excellent
particle handling capabilities of the multi N/C analyzers.
With the multi N/C 2300, this is achieved by the 0.7 mm
inner diameter of the aspiration needle in combination with
the ultra-short sample dosage line. The sample transfer
needles, tubing, and the rotary valve of the multi N/C 3300
with an inner diameter of 0.8 mm and, in combination with
the proven, fast sample loop injection principle, are perfectly
suited for samples containing particles. Additionally, the high
furnace temperatures guarantee 100% sample oxidation.
The use of large sample injection volumes and the routine
filling of the sample vessels further support a successful
cellulose particle test.
Real wastewater samples taken from the sewage treatment
process also show excellent reproducibility. A fully
Figure 3: multi NC 3300 with autosampler AS vario
Figure 2: Typical measurement curve and replicate results for the investigated cellulose suspension
automated measurement process for large sample sequences
is supported by a wide range of available autosamplers with
integrated sample homogenization, ranging from 21 sample
positions (AS 21hp) to 147 sample positions with the AS
vario and AS vario ER for the multi N/C 3300. The AS 60
autosampler for the multi N/C 2300 also offers a very good
level of automation for sewage plant testing laboratories